Patient Radiation Exposure Information
Patient radiation exposure information for common radiographic projections is presented in the table below. This information is available on form MIOSHA-RSS-880; Patient Radiation Exposure Information (PDF format). Clicking on an x-ray projection from the list will display the exposure distribution graph.
Projection (Click for Graph) |
Diagnostic Reference Level (mR) |
Achievable Level (mR) |
PA Chest (without a grid) | 15 | 10 |
PA Chest (with a grid) | 23 | 16 |
AP Lumbar Spine | 365 | 265 |
AP Cervical Spine | 77 | 55 |
DP Foot | 12 | 8 |
Hand | 7 | 5 |
Dental Cephalometric | 12 | 8 |
Dental Bitewing - all imaging modes | ||
- D Speed Film | 305 | 249 |
- E/F Speed Film | 180 | 135 |
- CR | 155 | 109 |
- DR | 115 | 83 |
Diagnostic Reference Level (mGy) |
Achievable Dose (mGy) |
|
CT Adult Head (CTDIvol) | 64 | 57 |
CT Pediatric Body (CTDIvol) | 12 | 8 |
LARA Dose Limit (mGy) |
||
Mammography Mean Glandular Dose | 3.0 | |
Cinefluorography Rates | ||
Dental Bitewing Imaging Modes Distribution |
Background |
Section health physicist inspectors routinely collect patient radiation exposure information during our regulatory inspections of x-ray facilities. The values measured are Exposure at Skin Entrance (ESE) measurements using technique factors used by the facility for an average-size patient. The exposure is measured free-in-air at the point where the x-rays enter the body. A chest phantom is used for photo-timed chests and the 16 cm CT phantom is used for CT adult head and pediatric body. |
Diagnostic Reference Levels |
Diagnostic Reference Levels (DRL) are based on data collected by the Radiation Safety Section during routine inspections of x-ray facilities. DRLs are set at the 75th percentile of the exposure distribution. Although DRLs are not regulations, they are considered by the Department to be upper limits which should not normally be exceeded for average-size patients. |
Achievable Levels |
Achievable levels are based on the median of the exposure distribution. The achievable level indicates a patient radiation exposure which should be readily attainable since 50% of Michigan facilities are already at this level. |
Exposure Distribution Graphs |
Graphs showing the distribution of radiation exposures over the preceding five years can be viewed by clicking on the desired x-ray projection listed below. These graphs are updated annually. They are used to review the Diagnostic Reference Levels and Achievable Levels. |
National NEXT Program |
The Nationwide Evaluation of X-Ray Trends (NEXT) is a national program conducted annually to measure the x-ray exposure that a standard patient receives for selected x-ray examinations. Click on this link for information on some of the NEXT surveys in the form of PDF trifolds. |