PBT Calibration Manual

Administrative Forms

PBT Instrument Calibration and Maintenance Log (TD-053)

TD-053 - Preliminary Breath Test Instrument Calibration and Maintenance Log

When a PBT is inspected, calibrated, or repaired, the TD-053 is used to document these actions.

  • Record all results obtained, regardless of whether they are within, or out of tolerance.

Completion and Retention:

  • Each PBT instrument shall have a separate log.
  • Begin a new log sheet each calendar year and retain the log for the current year plus three years.
  • Make an entry on the log sheet for each inspection and calibration.

Instructions for completion of the TD-053:

  1. ORI number of Department/Agency that owns the PBT instrument.
  2. Calendar Year - enter the current year.
  3. Department/Agency that owns the PBT instrument - printed department name.
  4. Instrument Serial Number - number located on the back of the PBT FC10.
  5. Instrument Model - enter model number (FC10).
  6. Date - month and day (i.e.: 5/20).
  7. Signature of operator - operator performing the maintenance, repair, or calibration.
  8. Operator's Department – print name of operator’s department.
  9. Operator ID Number- operator's certification number.
  10. Verification Results - results of the dry check (i.e., 0.078).
  11. Calibration/Maintenance Performed - print what operation was performed (“Verified” or "Calibrated").
  12. Dry Gas Lot Number - Enter the lot number appearing on the dry gas canister.